Tapered Pro Catalog & Manual

16 shop online at store.biohorizons.com SURGICAL KIT & DRILL SEQUENCE The implant driver section is color-coded by prosthetic platform (gray=3.0mm, yellow=3.5mm, green=4.5mm and blue=5.7mm). Abutment-level drivers are only for mounted implants. The surgical kit uses an intuitive layout to guide the surgeon through the instrument sequence. The sequence begins in the upper left hand corner and works left-to-right. Prior to use, clean and sterilize the surgical tray and instruments according to the Instructions for Use included with the kit. Study the surgical kit layout, color coding and iconography. Surgical assistants should be thoroughly familiar with all instruments and their uses prior to initiating the surgical procedure. Tapered HD drills increase in diameter as you work through the sequence from left-to-right.. Note: The 2.8mm drill is used for the 3.8mm diameter implant in soft bone, it is not needed for other sizes. soft bone dense bone 4.6mm body diameter 5.2mm body diameter soft bone dense bone 3.8mm body diameter soft bone dense bone 4.2mm body diameter soft bone dense bone HD Drill Sequence Implant Driver Recommended speed 1,500 - 2,000 RPM Recommended speed 1,000 RPM Maximum 30 RPM or use manually Initiate osteotomy Develop osteotomy Place implant matching the length of the prepared osteotomy