Tapered Short Catalog and Manual

10 shop online at www.biohorizons.com The osteotomy center-to-center measurement required to maintain a specific edge-to-edge spacing between two implants is calculated according to this formula: 1/2 (sum of 2 implant body diameters) + the desired spacing. The osteotomy centerpoint required to maintain a specific implantto-tooth spacing is calculated according to this formula: 1/2 (implant body diameter) + the desired spacing. During implant placement, clinicians must apply their best judgment as to the appropriate spacing for individual patient conditions. IMPLANT PLACEMENT LEVEL & SPACING osteotomy centerpoint edge-to-edge spacing When placing the implant in an uneven ridge, prepare the osteotomy and place the implant so the bone/soft-tissue junction is within the Laser-Lok transition zone. This will allow both soft tissue and bone to attach to the Laser-Lok collar. If the ridge discrepancy is more than the Laser-Lok transition zone, leveling the ridge can be considered. implant-totooth spacing osteotomy centerpoint