13 shop online at www.biohorizons.com FINAL BONE PREPARATION & PLACEMENT PEEK snap ring driver hex When seating the implant, use the corresponding dimples on the driver to orient one internal hex flat perpendicular to the implant angulation plane. Doing so verifies that an angled abutment will correct the angulation. Place the apex of the implant into the osteotomy and begin rotating slowly. The driver hex will engage when the driver is slowly rotated under apical pressure. If too much resistance is felt during insertion, reverse the implant to relieve pressure and re-insert into the osteotomy. If the final drill was not used while preparing the osteotomy, remove the implant and revise the osteotomy with the final drill. To pick up the implant, align the driver hex with the implant hex and press firmly to engage the PEEK snap ring. dimple hex flat Vial caps are color coded by body diameter (4.6mm=green, 5.8mm=blue). Cover caps and implant drivers are color coded by prosthetic platform (3.5mm=yellow, 4.5mm=green). Purpose: Engage the implant’s internal hex to drive implants into the osteotomy. • Implant level drivers are color coded by prosthetic connection: • yellow/green=3.5/4.5mm platform • 30 rpm or less4 Tapered Short implants can only be placed using HD drivers (TYGIDH & TYGIDR).