Tapered Short Guided Catalog and Manual

12 shop online at store.biohorizons.com Surgical Manual Guide Cylinders & Surgery Preparation Master Cylinders (pack of 10) Pre-surgery Preparation Inspect the surgical guide for defects and potential weak areas. Visually evaluate the position of the master cylinder to ensure it is placed according to the treatment plan. Ensure the thru-hole of the drill guides and master cylinders are free of debris. The surgical guide must have a stable fit to the patient’s anatomy. If a stable fit cannot be obtained at time of surgery, the surgical guide should not be used. Do not use excessive force to seat the surgical guide. Review the surgical plan and instruments within the kit prior to surgery. Drill use should be cross-checked against the drill usage chart (ML0232). Any drills that are worn, marked or dull should be replaced. Place the drill in the handpiece and check the fit with all the drill guides prior to surgery. TSGGMC-10 TSG 4.6mm Master Cylinder (pack of 10) TSGBMC-10 TSG 5.8mm Master Cylinder (pack of 10) Images: show the disassembled (screw and guide body) portions of the driver, vertically displaced from each other. Screw-retained Driver Assembly The screw-retained implant drivers are designed as a two piece assembly. If the part becomes disassembled during initial engagement of the implant or during removal from the implant after placement, please use the diagram to identify and reassemble the compatible parts. 6.0mm drivers of both diameters use the longer screw.