16 shop online at store.biohorizons.com The cover cap for a two-stage surgical protocol is mounted in the vial cap. Engage the implant with screw-retained driver by inserting the driver into the implant platform and tightening the screw. The driver can then be picked up with a handpiece by using the converter (CGS-4SC) or manually using a ratchet. Surgical Manual Implant Transfer & Guided Delivery Mount-free Transfer Vial caps are a surgical reference and are color-coded to indicate body diameter (4.6mm=green, 5.8mm=blue). Implant drivers are color-coded by prosthetic platform (3.5mm=yellow, 4.5mm=green) for proper mating with the implant connection. Length specific drivers are used in place of depth stops or stop position. Reference the patient-specific surgical protocol for the required implant driver. Guided Implant Delivery A handpiece or ratchet can be used to place the implant through the master cylinder. • Ensure the shaft of the implant driver is properly aligned with the master cylinder • Place the 5.8mm x 7.5mm implant through the master cylinder • If using a ratchet, an extender may be required to avoid collision with adjacent anatomy