BioHorizons Tapered HD surgical system

20 shop online at store.biohorizons.com SURGICAL KIT & DRILL SEQUENCE The implant driver section is color-coded by prosthetic platform (gray=3.0mm, yellow=3.5mm, green=4.5mm and blue=5.7mm). abutment-level drivers are only for mounted implants. The surgical kit uses an intuitive layout to guide the surgeon through the instrument sequence. The sequence begins in the upper left hand corner and works left-to-right and then down. Prior to use, clean and sterilize the surgical tray and instruments according to the Instructions for Use included with the kit. Study the surgical kit layout, color coding and iconography. Surgical assistants should be thoroughly familiar with all instruments and their uses prior to initiating the surgical procedure. The drilling section is color-coded by implant body diameter (gray=3.0mm, purple=3.4mm, yellow=3.8mm, green=4.6mm and blue=5.8mm). HD Drill Sequence HD Depth Gauge HD Crestal Bone Drill (site specific) HD Bone Tap (site specific) Implant Driver Recommended speed 1,500 - 2,000 RPM Recommended speed 1,000 RPM Recommended speed 1,000 RPM Maximum 30 RPM or use manually Maximum 30 RPM or use manually 3.8mm body diameter 3.4mm body diameter 3.0mm body diameter 4.6mm body diameter 5.8mm body diameter Note: The 2.8mm drill is used for the 3.4mm diameter implant, it is not needed for other sizes.

RkJQdWJsaXNoZXIy MTE0MzMw