BioHorizons TeethXPress® Manual

4 shop online at store.biohorizons.com SURGICAL PROTOCOL Initiate the anterior osteotomies using a BioHorizons Tapered Internal 2.0 diameter starter drill. Verify the correct implant trajectory using straight 2.0 parallel pins and the surgical guide. Note: In some instances, a 17°degree angled Multi-unit abutment may be necessary to create the proper angle through the guide trough. Following angled implant placement, screw in the BioHorizons profiler guide. Use a BioHorizons bone profiler to remove the crestal bone around the implant for a passive seating of the angled Multi-unit abutment. If the implant’s internal hex orientation is incorrect, use the BioHorizons implant-level ratchet driver to adjust the internal hex orientation to its ideal positioning for the angled Multi-unit abutment. Repeat the Multi-unit try-in abutment step to verify proper internal hex position. Use the BioHorizons 30° Multi-unit try-in abutment to verify proper implant hex orientation and to determine the necessary Multi-unit abutment height and angulation. Repeat steps on the opposing implant. Position the surgical guide to verify proper angulation and A/P spread. 5mm 4mm 3mm