Guided Surgery Kit Catalog and Manual

22 shop online at store.biohorizons.com GUIDED SURGERY CASE EXAMPLE Implant drivers include four stop positions for the depth stops to engage. Reference the patient-specific surgical protocol for the required stop position. Visual depth control can be used as an alternative to using the depth stops. Vial caps are a surgical reference and are color-coded to indicate body diameter (3.0mm=white, 3.4mm=purple, 3.8mm=yellow, 4.6mm=green, 5.8mm=blue). Implant drivers are color-coded by prosthetic platform (3.0mm=gray, 3.5mm=yellow, 4.5mm=green, 5.7mm=blue) for proper mating with the implant connection. The cover cap for a twostage surgical protocol is mounted in the vial cap. The laser-marked band above SP2 is used as a visual indicator to assist in delineating the four stop positions. stop position 3 (SP3) stop position 2 (SP2) stop position 1 (SP1) stop position 4 (SP4) Engage the implant with screw-retained driver by inserting the driver into the implant platform and tightening the screw. The driver can then be picked up with a handpiece by using the converter (CGS-4SC) or manually using a ratchet. The depth stop handle or disposable depth stops can be used for guided implant depth control. Disposable depth stops are for single use only.