Guided Surgery Kit Catalog and Manual

23 shop online at store.biohorizons.com GUIDED SURGERY CASE EXAMPLE A handpiece or ratchet can be used to place the implant through the master cylinder. If a pilot sleeve is being used, the surgical guide should be removed to deliver the implant using the conventional implant placement protocol. • Using the depth stop, engage the implant driver at the SP2 position • Ensure the shaft of the implant driver is properly aligned with the master cylinder • Place the 4.6 x 10.5mm implant through the master cylinder Place the implant through the master cylinder • Depth placement of the implant is controlled by the depth stop engaging the indicated stop position • The depth stop handle or disposable depth stop should firmly rest on top of the master cylinder Implant depth control Implant Placement depth stop handle disposable depth stop dimple dimple dimple dimple dimple dimple Orient the hex when using a ratchet, hand wrench or handpiece When seating the implant with a ratchet, use the corresponding dimples on the driver to orient one internal hex flat perpendicular to the implant angulation plane.